KingCreate Instrument IoT Module User Guide
KingCreate IoT Module Setup Guide for the KingCreate Communication Software
KingCreate IoT Module Setup Guide for Modbus Poll
STEP 1Connect to the Device Hotspot
- Power on the instrument and place your phone near the unit.
- Open your phone's Wi-Fi list and connect to the hotspot that begins with KC_.
- Open a browser and go to http://192.168.4.1.
STEP 2Configure Network Settings
- Configure the instrument Wi-Fi connection. Ensure the instrument and Modbus Poll are on the same LAN.
- Set Operating Mode to "Server Mode" and assign a static IP address that does not conflict with other devices.
STEP 3Save Settings
Click Submit to apply and save settings.
To re-enter setup mode, set the LOCK password to 128, power cycle the instrument, and then repeat Step 1.
STEP 4Configure Modbus Poll
- In Modbus Poll, select "Setup" > "Read/Write Definition."
- In the "Read/Write Definition" dialog, set the following:
- Slave ID: Instrument Modbus slave address (default: 1).
- Function: The required Modbus function code, for example, "Read Holding Registers" (03).
- Address: The starting register address, for example, 4097 (Channel 1 measured value). Refer to the instrument register map for details.
- Quantity: Number of registers to read (for example, 1), based on the data type and length.
- Click "OK" to save.
- Return to the main screen, then select "Connection" > "Connect."
- Type: Select "TCP/IP" as the connection type.
- IP Address: Instrument IP address (must match Step 2).
- Port: Instrument port number (must match Step 2).
- Response Timeout: 300 ms (default: 1000 ms).
- Delay between polls: 300 ms (default: 100 ms).
- Click "OK" to save.
- After the connection is established, Modbus Poll displays live instrument data for monitoring and analysis.